    The process viscometer PV100-TM64 works on the principle of a rotational viscometer
 
    with coaxial cylinder systeme.  
       
    Its basic components are a measuring head and a continuous-flow measuring unit  
    connected by a flange sleeve. The measuring head holds the drive unit, which is either a  
    controllable d.c. motor or a synchronous motor, as well as the torque sensor und the  
    electronics producing the measuring signal.  
       
    The rotation is transmitted from the measuring head over a connecting shaft to the  
    permanent-magnet coupling, which is located in the pressure-resistant measuring sleeve.  
    The magnetic field transfers the torque through the thin-walled measuring sleeve onto the  
    rotating measuring bob. This mesuring bob is enclosed by the measuring cup. In the  
    measuring sleeve and in the measuring cup, the measuring bob is held in low-friction pin  
    supports.  
       
    Measuring sleeve, measuring bob and measuring cup form a coaxial double gap cylinder  
    system.  
       
    Due to a specific design, part of the substance flow is continuously led through the
 
    measuring gap of the cylinder system. The viscosity-proportional torque occurring on the  
    measuring bob is measured in the measuring head and converted into an electric  
    standard current signal 4...20 mA, which can be used for display and as a control signal.  
       
    The immersion measuring device is designed for use with open containers or closed  
    vessels up to high operating pressures. The immersion depth of measuring system is  
    variable within wide limits by the length of the flange sleeve.  
       
    The process viscometer can be completed with an electronic unit for more functions of  
    instrument control, signal processing, as well as temperature adjustment of viscosity.  
       
 

Recommended Applications

 
    The process viscometer PV100-TM64 is suited for continuous viscosity measurement in a wide variety of substances, such as  
   
  • suspensions
     
       
  • diary products
     
       
  • lacquers or varnishes
     
       
  • printing inks
     
       
  • emulsion paints
     
       
  • lubricants
     
       
  • latex
     
       
  • polymer solutions
     
       
  • coal suspensions
     
       
  • glues or adhesives
     
       
  • resins
     
       
  • coating slips
     
       
  • chocolate suspensions
     
       
  • sealing compounds
     
       
  • fruit mash or preparations
     
       
  • vegetable mash
     
       
  • coating colours
     
       
  • cosmetics
     
       
  • solutions
     
       
  • emulsions
     
       
  • mud
     
        The rheology consultants of RheoTec will be pleased to help you analysing your specific application probleme, and, based on their profound rheological knowledge, they will find exactly the equipment configuration you need to solve it.  
           
      Technical data  
       
    Measuring Head MH10 MH16 MH16G
    viscosity range
    [mPas]
    2 to 10 4 20 to 10 5 10 3 to 10 7
    speed range [rpm]
    (8 adjustable steps)
    1 to 250 1 to 250 0.01 to 2.5
    in connection
    with spindle systeme
    MS-ZP1 MS-ZP2 MS-ZP2
    shear rate range
    [s -1 ]
    4.4 to 1.1 * 10 3 7.64 * 10 -1 to 191 7.64 * 10 -3 to 1.91
    shear stress range [Pa] 2.2 to 44 4 to 76.4 4 to 76.4
    accuracy 2.5 %
    pressure range up to 64 bar
    temperature range -20 ¡ÆC to +300 ¡ÆC
    mechanical connections flange connection DN80
    immersion depth 250 mm / 500 mm / 1,000 mm
    or 1,500 mm (other on request)
    materials for components that are exposed to the measured substance 1.4571 (DIN 17440)
    explosion protection II 2G EEx de IIB T6
    (according to ATEX standard)
    protection class IP54
    climatic conditions -25/55/30/90//42 (S0 2 )1-3
    installation position max. 30¡Æ deviation from vertical
    (or overhead operation)
    power supply motor:
    24 V DC, 500 mA
    signal circuit:
    24 V DC, 20 mA
    (supplied from an external source or by PV100 control unit VC1)
    measuring signal 4...20 mA to R a ¡Â 350 Ohm analogous, digital via RS485 interface
     
           
      Optional  
        1. PV100 control unit VC1  
       
     
       
    PV100 control unit VC1
     
        operations:  
       
  • supply of voltages 24 V DC for motor, signal circuit and temperature sensor
     
       
  • setting of viscosity range, speed step, limit value monitoring
     
       
  • measuring signal 4... 20 mA analog or digital via interface RS485 for viscosity and   temperature
     
       
  • data output on display (viscosity, measuring range, temperature, shear rate, shear   stress)
     
       
  • temperature correction of viscosity based on a fixed viscosity-temperature-equation   regarding a wide temperature range, display and output of reference viscosities
     
       
  • automatic range search run, overload protection
     
        The process viscometer electronics is available as plug-in rack in 19" systemes for one or two viscometers, or as splash-proof surface-type casing.  
           
        power supply : 90 to 253 V AC, 50/60 Hz  
        protection : IP20 (rack)
     
                    IP54 (surface-type casing)  
        2. Temperature sensor Pt100  
       
  • optional delivered with transducer
     
       
  • 4 temperature ranges: -30¡ÆC to +60¡ÆC, 0¡ÆC to +100¡ÆC, 0¡ÆC to +200¡ÆC, 0¡ÆC to +500¡ÆC
     
           
        3. Cooling unit KL  
        for reduction of heat conduction from probe to measuring head