Home>Á¡µµ°è/·¹¿À¸ÞŸ>Process Viscometer >Capillary Viscometer> KV100-Ex-DM25
 
 
Process Viscometer KV100-Ex-DM25
     
     
 
     

  À§ »çÁøÀ» Ŭ¸¯ÇϽøé Å©°Ô º¸½Ç ¼ö ÀÖ½À´Ï´Ù  
 
 
       
 
     
    The continuous-flow process viscometer KV100-Ex works on the principle of a Capillary  
    Viscometer with continual volume flow and measurement of pressure decrease over the
 
    measuring capillary.
 
       
    Its basic components are the driving head and the flow measuring device connected by a  
    spacer tube. The driving head holds a synchronous motor with change gear. The  
    immersion-type measuring unit holds a gear pump, the intake fitting and the exchangeable  
    measuring capillary, as well as ducts for the supply of the process pressure and for the  
    transport of overpressure from the entrance of the measuring capillary to the connections  
    of pressure transducers.
 
       
    The gear pump, which is driven by the driving head through a shaft via magnetic coupling  
    at constant speed, continually takes a constant volume flow from the process medium and  
    feeds it through the measuring capillary as a steady flow.  
       
    The pressure decrease between the two ends of the measuring capillary, which is  
    proportional to viscosity, is investigated by a differential pressure transducer. Pressure  
    differences are transformed into standard electrical signals 4...20 mA, which can be used  
    for display and as a control signal.  
       
    Screwed to the measuring device is a flow cell, which encloses the aspiration tube,  
    measuring capillary and pump housing. So these parts are directly in the process medium  
    positioned and temperature-controlled by this.  
       
    The flow measuring device is designed for use in the main or the secondary flow with low  
    operating pressures. The inlet of the aspiration tube is always located within the  
    representative substance flow.  
       
    The process viscometer can be completed with an electronics unit for more functions of  
    instrument control, signal processing, correction of characteristic lines of the gear pump,  
    as well as for the temperature adjustment of viscosity.  
       
 

Recommended Applications

 
    The process viscometer KV100-Ex-DM25 is suited for continual viscosity measurement and monitoring in a wide variety of liquid substances, such as  
   
  • lubricating oil
     
       
  • polymer solutions
     
       
  • fuel oil
     
       
  • emulsions
     
       
  • fat melts
     
       
  • suspensions
     
       
  • printing inks
     
       
  • liquid detergents
     
       
  • latex
     
       
  • adhesives
     
       
  • lacquers
     
       
  • glues
     
        The rheology consultants of RheoTec will be pleased to help you analysing your specific application problem, and, based on their profound rheological knowledge, they will find exactly the equipment configuration you need to solve it.  
           
      Technical data  
       
    viscosity ranges 0 to 3 mPas
    0 to 10 mPas
    0 to 30 mPas
    0 to 50 mPas
    0 to 100 mPas
    0 to 500 mPas
    accuracy ¡¾ 1 % of maximum value
    pressure range 0 to 2.5 MPa
    temperature range -25 ¡ÆC to +140 ¡ÆC
    materials with direct fluid contact 1.4571 according to DIN 17440
    explosion protection II 2G EEx de IIB T4
    (according to ATEX standard)
    degree of protection IP54
    climatic classification -25/55/30/90//42 (S0 2 )1-3
    installation position any
    power supply motor:
    230 V AC (+6% / -10% // 50 or 60 Hz)
    pressure difference transducer:
    24 V DC (+10% / -20% // 20 mA)
    data processing pressure difference transducer
    (¥Äp max. 0.05 MPa)
    output signal 4 to 20 mA (intrinsically safe)
     
           
      Optional Accessories  
      1. Continuous-flow cells for installation in main or secondary flow (bypass)
     
       
  • flow cell DMK (stainless steel) for pipeline installation (DN25...100, PN06/PN25)
     
       
  • flow cells DM06-K (synthetics) and DM25-E (stainless steel) for flexible tubes or small   pipe diameters
     
       
    flow cell DMK
    flow cell DM06
     
      2. Temperature sensor Pt100  
       
  • optional delivered with transducer
     
       
  • 3 temperature ranges -30¡ÆC to +60¡ÆC, 0¡ÆC to +100¡ÆC, 0¡ÆC to +200¡ÆC
     
      3. Control unit KS98  
       
  • provides supply voltage 24 V DC for differential pressure transducer
     
       
  • shows data on display (viscosity, temperature, min./max. viscosity, compensated   viscosity)
     
       
  • temperature correction of viscosity data according to specified viscosity-temperature   equation via large temperature range
     
       
    power supply: 90 to 250 V AC / 48 to 62 Hz
    protection: according to DIN VDE 0470 / EN 60529
    device panel: IP65
    housing: IP20
    connections: IP00
     
           
         
       
    Control Unit KS98