  The low temperature Viscometer has been designed to facilitate routine measurement of the viscosity of paints. Good correlation with application methods (e.g. brushing) is ensured by the use of a high rate of shear (10,000 sec-1). The cone & plate principle is selected because this is one of the most convenient ways of obtaining high rates of shear.  
       
  The ranges of measurement of 0-2, 0-5 poises and 0-10 poises are chosen as suitable for most normal paints. Alternative rates of shear or ranges of measurement can be obtained by using cones of alternative size or angle.  
       
  Temperature control is essential in measuring the viscosity of paint. This Viscometer has built-in provision for holding the sample at the temperature reagdless of normal variations in the bulk sample or in room temperature. Heating or cooling of the sample is by a Peltier effect semi-conductor device. The sample reaches the control temperature in no more than 15 seconds.
 
       
  "A sample a minute" is well within the capacity of this Viscometer.  
  The high temperature Viscometer has been designed for the control of the viscosity of resins, bitumens, hot melts and similar products, during manufacture.  
       
  These Viscometers enable a very small sample from the cooking vessel to be checked for viscosity in less than a minute. The instrument is thermostatically controlled over a wide range of temperatures, one of which is selected so that the melt viscosity of the resin comes within the viscosity of the instrument.  
       
  In this way, thinning tests are eliminated and a quick and accurate determination of resin viscosity is obtained.  
     
  Ease of cleaning is ensured by virtue of the small sample required and the ease of access which is possible with the simple geometry of the cone and plate principle. A wipe with paper tissue and suitable solvent is all that is normally required.
 
           
  Specifications:
 
   
  Electrical supply   220-240 Volts or 110-120 Volts 50/60 Hz
  single phase.
  Power   125 Watts
  Standard viscosity range   0-2 poises at 10,000 sec-1
  0-2.5 poises at 10,000 sec-1
  0-5 poises at 10,000 sec-1
  0-10 poises at 10,000 sec-1
  0-20 poises at 2,500 sec-1
  0-40 poises at 2500 sec-1
  0-100 poises at 2500 sec-1
  Cone Type   To suit range: 0-2, 0-2.5, 0-5, 0-10,
  0-20,0-40, 0-100 poises
  Accuracy   +/- 2% of full scale, confirmed using
   PRAoils traceable to National    Standards.
  Warm-up from cold   15 minutes
  Temperature control of sample   One or three switched temperatures in
  the range 10 degrees C - 75 degrees C,
  six temperatures in the range 25   degrees C - 230 degrees C.
  Time for sample to obtain control temperature   20 seconds
  Ambient Temperature H.T. Instrument   To calibrate the high temperature
  instrument at 25 degrees C, the ambient
  temperature should be <20 degrees C.
  Ambient Temperature L.T. Instrument   The ambient temperature will affect the
  instrument lowest temperature. Lowest
  temperature achieved is 10 degrees C.
  below ambient.